Procedure for Bringing LENA Back Up After Safing

 

 

 

All verifiers appear on Tab "HK" unless otherwise indicated.

 

 

 

 1. /LENA_ON RT

 

        a. About 10 minutes after command sent, "Program Mode" should change to "RAM" (Window: Normal HK)

        b. "Software Version" should change to 36 (for 3.6) (Window: Normal HK)

        c. About 6 minutes later, the "Autonomous Test" should become disabled (Tab: Mem; Window: norma_HK)

        d. Wait for update - command counter to 0 (Window: Normal HK)

        e. Wait for Built-in-test = "Off" (Window: Normal HK)

 

 

 2. /L_HVP_GBL_USAFE RT

 

        a. HVPS Safe should be disabled (Window: Nom 2-HK)

        b. Wait for update - command counter to 1 (Window: Normal HK)

 

 

 3. /L_SYS_HVP_VTHR  MCPSTRT=x'AA' MCPSTP=x'AA' COLL_P=x'74' COLL_N=x'74' OPTICS=x'51' RT

 

 

 4. /L_SYS_HVP_ITHR  MCPSTRT=x'9E' MCPSTP=x'8B' COLL_P=x'FF' COLL_N=x'C0' OPTICS=x'6E' RT

 

        a. These changes in threshold should be reflected on HK screen - Final currents

should read 12.4, 10.9, 20, 15.1, 8.62 mA

        b. Final threshold voltage values should read 2, 2, 4, 4, 6.99 kV (Window: diagnos)

        c. Wait for update - command counter to 3 (Window: Normal HK)

 

 

 5. /L_HVP_MSTP_CMD V=x'00' RT

 

 

 6. /L_HVP_MSTR_CMD V=x'00' RT

 

        a. Since the initialized values are probably zero, look for the command counter to

go to 5 (Window: Normal HK)

        b. Wait for update

 

 

 7. /L_SYS_OVCT_THR  P1=x'09' RT

 

        a. Look for TOF overcount threshold to go to 9 (from default value of 65535) (Window: norma_hk)

        b. Wait for update - command counter to 6 (Window: Normal HK)

 

 

 8. /L_HVP_MSTR_EN RT

 

 

 9. /L_HVP_MSTP_EN RT

 

 

10. /L_HVP_MSTR_CMD  V=x'40' RT

 

 

11. /L_HVP_MSTP_CMD  V=x'40' RT

 

        a. Start and Stop MCP power supplies should be enabled (Window: Nom2-HK)

        b. Start and Stop MCP levels should go to 40 hex (1140 V/1140 V on anode)

        c. Wait for update - command counter to 10

 

 

12. /L_HVP_MSTR_CMD  V=x'80' RT

 

 

13. /L_HVP_MSTP_CMD  V=x'80' RT

 

 

14. /L_SYS_HVP_VTHR  MCPSTRT=x'D5' MCPSTP=x'D5' COLL_P=x'91' COLL_N=x'91' OPTICS=x'73' RT

 

        a. Start and Stop MCP levels should go to 80 hex (1.8 kV/1.8 kV on anode) (Window: Nom2-HK)

        b. Final threshold voltage levels should read 2.5, 2.5, 5, 5, 9.92 kV (Window: diagnos)

        c. Wait for update - command counter to 13 (Window: Normal HK)

 

 

15. /L_HVP_MSTR_CMD  V=x'A0' RT

 

 

16. /L_HVP_MSTP_CMD  V=x'90' RT

 

        a. Old nominal levels

        b. Should be getting counts on Start and Stop MCPs (Tab: Singles)

        c. Start and Stop MCP levels should go to A0 hex and 90 hex (2.1 kV/1.95 kV on

anode) (Window: Nom2-HK)

        d. Wait for update - command counter to 15 (Window: Normal HK)

 

 

17. /L_HVP_OPT_EN RT

 

        a. Optics should be enabled (Window: Nom2-HK)

        b. Optics should read about 2 kV (Window: Nom2-HK)

        c. Wait for update - command counter to 16 (Window: Normal HK)

 

 

18. /L_HVP_COLN_EN RT

 

        a. Negative collimator should be enabled (Window: Nom2-HK)

        b. Negative collimator should read 1.8 kV (Window: Nom2-HK)

        c. Wait for update - command counter to 17 (Window: Normal HK)

 

 

19. /L_HVP_OPT_CMD  V=x'20' RT

 

        a. Optics level should go to 20 hex (Window: Nom2-HK)

        b. Optics voltage should go to 4.51 kV (Window: Nom2-HK)

        c. Optics current should be about 4.4 mA (Tab: HVPS Strip Charts)

        d. Wait for update - command counter to 18 (Window: Normal HK)

 

 

20. /L_HVP_OPT_CMD  V=x'40' RT

 

        a. Optics level should go to 40 hex (Window: Nom2-HK)

        b. Optics voltage should go to 7.02 kV (Window: Nom2-HK)

        c. Optics current should be about 6.0 mA (Tab: HVPS Strip Charts)

        d. Wait for update - command counter to 19 (Window: Normal HK)

 

 

21. /L_HVP_OPT_CMD  V=x'50' RT

 

        a. Optics level should go to 50 hex (Window: Nom2-HK)

        b. Optics voltage should go to 8.27 kV (Window: Nom2-HK)

        c. Optics current should be about 6.2 mA (Tab: HVPS Strip Charts)

        d. Wait for update - command counter to 20 (Window: Normal HK)

 

 

22. /L_HVP_OPT_CMD  V=x'60' RT

 

 

23. /L_SYS_HVP_VTHR MCPSTRT=x'D5' MCPSTP=x'D5' COLL_P=x'91' COLL_N=x'91' OPTICS=x'97' RT

 

        a. Optics level should go to 60 hex (Window: Nom2-HK)

        b. Optics voltage should be about 9.53 kV (Window: Nom2-HK)

        c. Optics current should be about 6.8 mA (Tab: HVPS Strip Charts)

        d. Final threshold voltage values should read 2.5, 2.5, 5, 5, 13 kV (Window: diagnos)

        e. Wait for update - command counter to 22 (Window: Normal HK)

        f. Wait at least 5 minutes before sending the next optics command

 

 

24. /L_TOF_TRSH_START  CFDTRSH=10 RT

 

        a. Start CFD threshold should go to 10 decimal (Window: Normal HK)

        b. Wait for update - command counter to 23 (Window: Normal HK)

 

 

25. /L_HVP_OPT_CMD  V=x'70' RT

 

        a. Optics level should go to 70 hex (Window: Nom2-HK)

        b. Optics voltage should be about 10.8 kV (Window: Nom2-HK)

        c. Optics current should be about 7.53 mA (Tab: HVPS Strip Charts)

        d. Wait for update - command counter to 24 (Window: Normal HK)

        e. Wait at least 5 minutes before sending the next optics command

 

 

26. /L_ROI_CNFG_PKG0  ROIDEF=[x'3F',x'8',x'0B',x'2',x'0',x'2C'] RT

 

 

27. /L_SYS_ROI_SEL  ROISEG=x'01' RT

 

        a. ROI0 should be enabled (Window: Nom2-HK)

        b. Wait for update - command counter to 26 (Window: Normal HK)

        c. "Old" nominal level reached here

 

 

28. /L_HVP_OPT_CMD  V=x'78' RT

 

        a. Optics level should go to 78 hex (Window: Nom2-HK)

        b. Optics voltage should be about 11.4 kV (Window: Nom2-HK)

        c. Optics current should be about 7.92 mA (Tab: HVPS Strip Charts)

        d. Wait for update - command counter to 27 (Window: Normal HK)

        e. Monitor instrument performance for at least 5 minutes to assure normalcy

 

 

29. /L_SYS_STAT_SAVE RT

 

        a. Wait for update - command counter to 28 (Window: Normal HK)

 

 

30. /L_HVP_MSTR_CMD  V=x'AA' RT

 

        a. Start MCP level should go to AA hex (2.2 kV on anode) (Window: Nom2-HK)

        b. Wait for update - command counter to 29 (Window: Normal HK)

 

 

31. /L_HVP_MSTR_CMD  V=x'BF' RT

 

        a. Start MCP level should go to BF hex (2.4 kV on anode) (Window: Nom2-HK)

        b. Wait for update - command counter to 30 (Window: Normal HK)

 

 

32. /L_HVP_MSTP_CMD V=x'95' RT

 

        a. Stop MCP level should go to 95 hex (2 kV on anode) (Window: Nom2-HK)

        b. Wait for update - command counter to 31 (Window: Normal HK)

 

 

33. /L_SYS_STAT_SAVE RT

 

        a. Wait for update - command counter to 32 (Window: Normal HK)

 

 

34. /L_SYS_RSTR_EN RT

 

        a. Autonomous State Restore should be enabled (Tab: Mem; Window: norma_hk)

        b. On Monitor screen, MCP modulate should be enabled (Tab: Monitor; Window: monitor_hkv2)

        c. Wait for update - command counter to 33 (Window: Normal HK)